Permit Information
Permit NumberELE06-00179
StatusFinaled
Name UNITED ELECTRIC CONTRACTING
Site Address 8800 W EMERALD ST
Project NameHEALTHSOUTH TVH
Customer Number 
ePlanReview 
Description
Remodel MRI suite and hook up new power, misc outlets, power alarm with misc wiring and install floor duct per phone call from jenny with United ele initial value of $56,000///1/17/06 BAS
Contact Information
See Activities list for contact information
Show All Activities
DescriptionReceivedRequestedCompletedAM/PMDispositionAssigned To 
Final7/27/20067/28/20067/28/2006APASS
DetailsBob Reichert
Phone #: ###-#### Internet Inspection Request from: JENNY EMEL corr made, ok to final
Final7/10/20067/11/20067/11/2006AFAIL
DetailsBob Reichert
PLEASE CALL DON AT ###-#### 30 MIN BEFORE ARRIVAL TO MEET INSPECTOR; Phone #: ###-#### Internet Inspection Request from: JENNY EMEL BCC-4-05-33(A)(B) all electrical equipment shall be listed and labelled by a nationally recognized testing laboratory as an assembly, nec 400.8(1) flexible cords and cables shall not be used as a substitute for the fixed wiring, Note: corrections shall be made within 10 days or further action will be taken
Final7/10/20067/11/20067/10/2006ACANC
DetailsBob Reichert
FINAL FOR THE MRI REMODEL; Phone #: ###-#### Internet Inspection Request from: JENNY EMEL Inspection cancelled from web by: JENNY EMEL
ELE-Aging Inspection5/11/20065/11/20065/11/2006 DONE
DetailsBob Reichert
talked to contr,suc
Electrical Rough-In3/10/20063/10/20063/10/2006PPASS
DetailsBob Reichert
ceiling ok to cover mri room
Electrical Rough-In2/15/20062/16/20062/16/2006APASS
DetailsBob Reichert
Phone #: ###-#### Internet Inspection Request from: JENNY EMEL walls only for mri room and control room ok to cover
Electrical Rough-In1/17/20061/18/20061/18/2006APASS
DetailsBob Reichert
rick ###-#### file room walls and ceiling ok to cover