Final | 6/14/2015 | 6/15/2015 | 6/15/2015 | A | PASS |
Daryl DeGrange
|
|
|
Phone #: ###-#### Inspection requested via web by Cyndi Dale X-ray equipment performance evaluation completed Medical Imaging, Inc. final ok
|
Final | 6/3/2015 | 6/4/2015 | 6/4/2015 | A | PART |
Daryl DeGrange
|
|
|
Phone #: ###-#### Inspection requested via web by Cyndi Dale Temporary occupancy only till corrections on x-ray equipment completed
|
Final | 6/1/2015 | 6/2/2015 | 6/3/2015 | A | FAIL |
Daryl DeGrange
|
|
|
please contact cory with questions. ###-#### Phone #: ###-#### Inspection requested via web by Cyndi Dale Connecting cabling entering enclosurers without connectors nor abrasion protection, exposed termimals electrical slices
|
Final | 6/1/2015 | 6/1/2015 | 6/1/2015 | P | FAIL |
Rex Boyles
|
|
|
Please call Cory with any questions. ###-#### thanks, Phone #: ###-#### Inspection requested via web by Cyndi Dale 517.16- Receptacles in exam rooms need to be "hospital grade". The receptacles installed are "isolated ground".
|
Final | 5/19/2015 | 5/20/2015 | 5/20/2015 | A | FAIL |
Rex Boyles
|
|
|
Phone #: ###-#### Inspection requested via web by Cyndi Dale as per article 100 in 2014 NEC equipment must have label for UL listing (XRAY)
|
Electrical Rough-In | 4/17/2015 | 4/17/2015 | 4/17/2015 | P | PASS |
Rex Boyles
|
|
|
reinspection for walls. - right pillar in front of dive magic - there is a key lock box. code is 4927 Phone #: ###-#### Inspection requested via web by Cyndi Dale corrections made r/i walls ok
|
Electrical Rough-In | 4/15/2015 | 4/16/2015 | 4/16/2015 | A | FAIL |
Rex Boyles
|
|
|
Wall Inspection. right pillar in front of dive magic - there is a key lock box. code is 4927 Phone #: ###-#### Inspection requested via web by Cyndi Dale 517.13(B)-MC cable must be redundant type for patient care areas. See 517 definitions and general care space.
|