Description |
ST LUKES - IE CT SCAN - 1ST FLR) To remodel an existing 523 square foot room for upgrade and installation of a new CT scanner. Selective interior of room demolition is to occur; lead lining & wall cover are to be added at interior room walls; structural plate & anchors to be installed for boom support at bottom of existing 2-hour concrete floor assembly; add 2 new doors, a new 45 minute door assembly/frame/hardware at 1-hour fire barrier penetration and a 90 minute door/frame/hardware assembly at door opening into the existing 2-hour exit passageway stair enclosure extension to the exterior; maintain and repair the existing 2-hour mechanical shafts and protect penetrations; sprinkler & alarm systems to meet tenant needs and comply with current adopted codes and ordinances. This building is fully sprinklered for occupancy use group compliance. The sprinkler work scope consists of dropping 4 pendant heads into a new wall and ceiling layout. This permit is attached to BLD06-02302.
|