Final | 11/4/2003 | 11/4/2003 | 11/4/2003 | P | PASS |
Lee Rice
|
|
|
Phone #: ###-#### Internet Inspection Request from: Sonja Hankel
|
Final | 10/29/2003 | 10/30/2003 | 10/30/2003 | P | PART |
Jim Johnson
|
|
|
need to finish installation of all equipment 10/30/2003 2:20:29 PM
|
Final | 10/28/2003 | 10/29/2003 | 10/29/2003 | P | NR |
Jim Johnson
|
|
|
Phone #: ###-#### Internet Inspection Request from: Sonja Hankel not ready 10/29/2003 3:20:22 PM
|
Plumbing Rough-in | 9/12/2003 | 9/12/2003 | 9/12/2003 | A | PASS |
Jim Johnson
|
|
|
|
Med/Gas Systems | 9/10/2003 | 9/11/2003 | 9/11/2003 | A | FAIL |
Jim Johnson
|
|
|
test on vacuum system failed also need to correct vent location on island sink vent 9/11/2003 1:01:19 PM
|
Plumbing Rough-in | 9/9/2003 | 9/10/2003 | 9/10/2003 | A | PART |
Jim Johnson
|
|
|
check med gas inspection
thank you; Phone #: ###-#### Internet Inspection Request from: Liz Sasser start of 24 hr. med gas pressure test on rough- in plumbing need to correct vent on island sink , regrade and support drain line to island sink correct leak in drainage y- fitting in basement over head ( note received and oked revised plumbing plan on 9/5/03 ) 9/10/2003 12:13:11 PM
|
Plumbing Rough-in | 9/8/2003 | 9/8/2003 | 9/8/2003 | P | FAIL |
Jim Johnson
|
|
|
inspection late pm; Phone #: ###-#### Internet Inspection Request from: Liz Sasser no access building locked up 9/8/2003 4:47:54 PM
|
Plumbing Rough-in | 8/27/2003 | 8/27/2003 | 8/27/2003 | P | FAIL |
Jim Johnson
|
|
|
Phone #: ###-#### Internet Inspection Request from: Liz Sasser repipe island sink,s as per upc 909.0 need additional cleanouts in trap arm,s exceeding 90 as per upc 1002.3 need to support all piping as per table 3-2 need to correct all venting that exceed 1/3 of permitted length as per table 7-5 need to submitt plan showing all additional plumbing in operatory room,s and relocation of med gas room 8/27/2003 5:00:34 PM
|
Plumbing Rough-in | 8/26/2003 | 8/26/2003 | 8/26/2003 | P | CANC |
Jim Johnson
|
|
|
Phone #: ###-#### Internet Inspection Request from: Liz Sasser Inspection cancelled from web by: Liz Sasser
|
Groundwork | 6/27/2003 | 6/27/2003 | 6/27/2003 | P | PASS |
Jim Johnson
|
|
|
###-#### debest
|
Water and/or Sewer | 6/27/2003 | 6/27/2003 | 6/27/2003 | P | PASS |
Jim Johnson
|
|
|
###-#### debest
|
Water and/or Sewer | 6/26/2003 | 6/26/2003 | 6/26/2003 | P | PART |
Jim Johnson
|
|
|
###-#### OFFICE UNDER GROUND SWR & WATER sewer line test ok ok to cover need to change water line to 1 1/2 as per plans 6/26/2003 3:48:40 PM
|