Third Notice ELE | 8/22/2005 | 8/22/2005 | 8/29/2005 | | DONE |
|
|
|
Second Notice has not been responded to.
|
Second Notice ELE | 8/9/2005 | 8/9/2005 | 8/22/2005 | | DONE |
|
|
|
First Notice has not been responded to.
|
First Notice | 7/27/2005 | 7/27/2005 | 8/9/2005 | | DONE |
Jack Frank
|
|
|
|
Final | 7/25/2005 | 7/27/2005 | 7/27/2005 | P | PASS |
Jack Frank
|
|
|
mica; Phone #: ###-#### Internet Inspection Request from: Letitia Mesa ok
|
Final | 6/13/2005 | 6/14/2005 | 6/14/2005 | P | FAIL |
Jack Frank
|
|
|
Scheduled by a contractor via Hello! on 6/13/2005 at 2:04:18 PM. Contact: ###-####. called shall not use flex 517.78(C) grding shall meet 517.13(A&B)
|
Final | 6/7/2005 | 6/8/2005 | 6/8/2005 | A | PART |
Jack Frank
|
|
|
Scheduled by a contractor via Hello! on 6/7/2005 at 1:05:42 PM. Contact: ###-####. T/O only stills needs testing of x ray equip, which is set up for next week with UL
|
Final | 5/26/2005 | 5/26/2005 | 5/26/2005 | P | FAIL |
Jack Frank
|
|
|
Scheduled by a contractor via Hello! on 5/26/2005 at 8:58:16 AM. Contact: ###-####. no final structural cor posted states needs shell and core final, before tenant space can be finaled, also 250.119 shall not ID 6 or smaller grd xray equip, 110.2&5
|
Electrical Rough-In | 3/23/2005 | 3/24/2005 | 3/24/2005 | A | PASS |
Bob Reichert
|
|
|
Scheduled by a contractor via Hello! on 3/23/2005 at 1:32:06 PM. changed panel location, provided pipng for x-ray, and redundant gndg same. ric ok .elect was there and reviewed changes. 3/24/2005 8:23:16 AM
|
Electrical Rough-In | 3/22/2005 | 3/22/2005 | 3/22/2005 | P | PART |
Jack Frank
|
|
|
Scheduled by a contractor via Hello! on 3/22/2005 at 7:34:58 AM. Contact: ###-####. walls ok needs to address wiring and xray equip. also noted listing for xray 3/22/2005 1:29:43 PM
|
Application received | | | 3/18/2005 | | |
|
|
|
|
Issue permit | | | 3/18/2005 | | DONE |
|
|
|
|