Case Summary (ELE) | | | 2/18/2016 | | DONE |
|
|
|
|
First Notice | 10/3/2013 | 10/3/2013 | 11/26/2013 | | DONE |
Daryl DeGrange
|
|
|
|
Final | 10/2/2013 | 10/3/2013 | 10/3/2013 | A | PASS |
Daryl DeGrange
|
|
|
Please call Lynn at ###-#### with any questions comments or concerns. Phone #: ###-#### >> Inspection requested via web by Lynn Proctor final ok
|
Electrical Rough-In | 9/26/2013 | 9/27/2013 | 9/27/2013 | P | PASS |
Daryl DeGrange
|
|
|
Please call Lynn at ###-#### with any questions, comments or concerns. Phone #: ###-#### >> Inspection requested via web by Lynn Proctor rough-in ok cover ceiling grid
|
Final | 9/26/2013 | 9/27/2013 | 9/26/2013 | P | CANC |
Daryl DeGrange
|
|
|
Please call Lynn at ###-#### with any questions, comments or concerns. Phone #: ###-#### >> Inspection requested via web by Lynn Proctor Inspection cancelled from web by: Lynn Proctor
|
Electrical Rough-In | 9/25/2013 | 9/26/2013 | 9/26/2013 | A | NR |
Daryl DeGrange
|
|
|
Please call Lynn with any questions, comments or concerns. Phone #: ###-#### >> Inspection requested via web by Lynn Proctor Not ready
|
Electrical Rough-In | 9/24/2013 | 9/25/2013 | 9/25/2013 | P | FAIL |
Bob Reichert
|
|
|
Phone #: ###-#### >> Inspection requested via web by Lynn Proctor equipment shall be listed, BCC 4-05 -33(A)(B) shall not use so cord for permanent wiring, NEC 400, shall provide seismic wires for filters in ceiling, BCC
|
Electrical Rough-In | 9/10/2013 | 9/11/2013 | 9/11/2013 | P | PASS |
Daryl DeGrange
|
|
|
Bud on site. Wall rough in for PET/CT Scan Phone #: ###-#### >> Inspection requested via web by Clint Tate rough-in ok cover lab ceiling grid and east wall
|
Issue PAID IN FULL permit | | | 9/10/2013 | | DONE |
|
|
|
Payment received via Govolutions
|
Application received | | | 9/10/2013 | | DONE |
|
|
|
This permit was created on the web by Lynn Proctor
|