First Notice | 10/7/2015 | 10/7/2015 | 12/18/2015 | | DONE |
Rock A. Hudson
|
|
|
This permit finaled on the web by Brad Baker
|
Print Statement | | | 12/7/2015 | | |
|
|
|
|
Final | 10/7/2015 | 10/7/2015 | 10/7/2015 | A | PASS |
Rock A. Hudson
|
|
|
Corrections have been taken care of please call me this morning with a email address so I can send you the xray data Phone #: ###-#### Inspection requested via web by Brad Baker Final ok
|
Final | 10/5/2015 | 10/6/2015 | 10/6/2015 | A | FAIL |
Rock A. Hudson
|
|
|
Code to door if needed is 9500 Phone #: ###-#### Inspection requested via web by Brad Baker 210.8(B)(5)- All 15 and 20 amp 120 volt receptacle outlets located with in 6' of a sink shall have GFCI protection. 110.3(B)- Confirm listing of x-ray unit and controller as listed system. 300.4(G)- Conduits with #4 or larger conductor shall have a bushing to protect conductors. 330.30(B)- MC cable shall be secured within 12" of panel. * Several receptacle outlets in offices not working.
|
Electrical Rough-In | 9/18/2015 | 9/18/2015 | 9/18/2015 | P | PASS |
Daryl DeGrange
|
|
|
It should be open and ready for inspection on walls and ceilings for cover Phone #: ###-#### Inspection requested via web by Brad Baker phone and data cabling removed do to 300.4 violation rough-in ok cover interior walls and hard lid areas
|
Issue PAID IN FULL permit | | | 9/17/2015 | | DONE |
|
|
|
Payment received via Govolutions
|
Owner Supplied Equipment | | | 9/17/2015 | | |
|
|
|
|
Application received | | | 9/17/2015 | | DONE |
|
|
|
This permit was created on the web by Brad Baker
|