Water and/or Sewer | 3/2/2001 | 3/2/2001 | 3/2/2001 | A | PASS |
Bill VanDegrift
|
|
|
Added by Inspector: 330 On 3/2/2001 11:36:04 AM
|
DOC - Permit Finaled | 3/2/2001 | 3/2/2001 | 3/2/2001 | A | PASS |
Bill VanDegrift
|
|
|
Added by Inspector: 330 On 3/2/2001 11:36:22 AM
|
Final | 2/23/2001 | 2/26/2001 | 2/26/2001 | A | PART |
Bill VanDegrift
|
|
|
Inselman Plumbing ###-#### med gas cert ok 90's on discharge ok still needs hangers on vacuum discharge piping in ceiling and cap equipment drain no longer used in dark room 2/26/2001 11:09:32 AM
|
Final | 2/19/2001 | 2/20/2001 | 2/20/2001 | A | PART |
Bill VanDegrift
|
|
|
Scheduled by a contractor via Hello! on 02/19/2001 at 15:13:33. Contact: ###-####. provide a copy of med gas cert, hang vacuum discharge piping and turn down above roof, temp occ ok, ok to occupy but no surgery prior to med gas cert. 2/20/2001 10:38:23 AM
|
Plumbing Rough-in | 12/18/2000 | 12/18/2000 | 12/18/2000 | P | PASS |
Lee Rice
|
|
|
Scheduled by a contractor via Hello! on 12/18/2000 at 10:20:36. Contact: ###-####.med gas
|
Application received | | | 12/18/2000 | | |
|
|
|
|
Issue & PRINT permit | | | 12/18/2000 | | DONE |
|
|
|
|